Post-Installation Control

5–10
INSTALLING THE CEPHALOSTAT
Post-Installation Control
Before the post-installation control, check that:
The installation of the unit is complete.
The installation of the KODAK dental imaging software is complete.
You have the test tool.
You have raised the digital sensor to ceph mode.
The unit is on.

Adjusting the Cephalostat Head

Adjusting the Cephalostat Head Inclination

To adjust the cephalostat head inclination, follow these steps:
1. Position the head clamps in a frontal AP position.
2. Access the Cephalometric Acquisition Window. In the Program pane, click
3. Launch an x-ray using the x-ray remote control. The head clamps must be correctly
and vertically positioned on the acquired x-ray image.
.