Electrical Requirements - Kodak 8000C Installation Manual

Digital panoramic and cephalometric extraoral imaging system
Hide thumbs Also See for 8000C:
Table of Contents

Advertisement

Electrical Requirements

WARNING*
You MUST select the operating voltage when placing an order. The
operating voltage CANNOT be modified on site.
The unit can operate at:
100/110/130 V 50/60 Hz
230/240 V 50/60 Hz
Table 4–1
Optional Operating Voltages of the Unit
Nominal voltage *
(no load)
100 V - 130 V
230 V - 240 V
CAUTION
The power supply line must be equipped with a connection box that
ensures a constant connection. It must not be possible to connect the
unit to the power supply without using a tool. The unit must be
protected against any accidental disconnection.
If other units are installed on the same line, interference and voltage fluctuations can
cause the radiological unit to operate abnormally. We strongly recommend that a
separate electrical line be dedicated to supply power to the KODAK 8000C unit.
This line should be protected by a circuit breaker with a maximum current of:
16 A at 230/240 V
20 A at 110/130 V
A differential circuit breaker of 30 mA
KODAK 8000C Digital Panoramic and Cephalometric Extraoral Imaging System Installation Guide (SM736)_Ed 02
Minimum
Maximum
90 V
143 V
207 V
264 V
Electrical Requirements
Maximum line
current
20 A
10 A
4–3