Connecting the Exhaust Hose
The building ventilation system must
draw air to the outside of the building,
so that no air is reused.
[1]
Connect a 7.6 cm (3 inch) EXHAUST HOSE,
not supplied, to the EXHAUST ADAPTER.
[2]
Connect the EXHAUST HOSE to the
EXHAUST DUCT from the building ventilation
system.
[3]
Leave a 5 cm (2 inch) air gap between the
EXHAUST HOSE and the EXHAUST DUCT.
• If the venting is not correct, fumes will corrode equipment and cause artifacts. Do not install the
processor or accessories if the venting is not correct.
• The building ventilation system must draw air to the outside of the building, so that no air is reused.
• The AIR FLOW is correct when the fumes are flowing out of the processor through the EXHAUST
HOSE. Before installing the processor, or at the next service call, do the following to check that the
AIR FLOW is correct.
EXHAUST ADAPTER
981159
Environmental Requirements
IMPORTANT
EXHAUST
DUCT
EXHAUST
HOSE
5 cm (2 in.)
maximum
Clearance for
Exhaust Duct