6.9

System Reset

1. Unplug AC and the AC LED that is on the base panel should be off, and vice versa.
2. While the AC is plugged in, turn on hand pendant, and its panel should be lit.
3. The battery power indicator should come on after the hand pendant is turned on.
4. While the hand pendant is on, turn the (C) knob so the indicator is pointing straight down at
the 6 o'clock position.
5. Flip the (B) toggle switch all the way down.
6. Press and hold the (A) toggle switch down for 5
seconds until the hand pendant is turned off,
which means the system has been reset. While
the switch is being pressed down, the pump motor
will be running.
7. Return the (B) toggle switch to its neutral position
(in the middle).
Figure 6-3
6-10
B2AV800 Series Surgical Table
IM-B2AV800-E01