4.6.1 Performance Checklist

After completing installation procedures (4.1 to 4.5), installation technician shall verify the
performance of table (also refer to 5.2.1 "Hand Pendant Operation") according to the specifications
and record the results of verification in the chart below:
Test Item
Height range of table
without mattress pads
Slide
Kidney Elevator
Trendelenburg
Reverse Trendelenburg
Lateral Tilt
Back Section
Leg Section
Head Section
Flex
Reflex
Override Switches
Level
Specification
26.8 in ~ 44 in
(68 cm ~ 112 cm)
12.2 in (31cm)
4.7 in (12 cm)
0° ~ 30°
0° ~ 30°
to right 0° ~ 20°
to left 0° ~ 20°
up 0°~80°
o
down 0°~50
up 0° ~ 30°
down 0° ~ 90°
left and right 180°
up 0° ~ 90°
down 0° ~ 45°
Back section down 25° max
Seat section down 25° max
Back section up 30° max
Seat section up 30° max
Height, Trendelenberg, reverse
Trendelenberg, side tilt, back
section, slide (for
B2AV800LSK, LS, SK, S) floor
locks (for B2AV800 LSK, LS)
+/- 2°
Result
4-6
B2AV800 Series Surgical Table
IM-B2AV800-E01
Verified