GENERAL INFORMATION

3.1 Product Description

There are six models in the B2AV800 Series as shown in the following table:
Table 3-1: Model-Function Comparison
Function
Model
B2AV800-1 (-2)
B2AV800K-1 (-2)
B2AV800S-1 (-2)
B2AV800SK-1 (-2)
B2AV800L-1 (-2)
B2AV800LK-1 (-2)
B2AV800LS-1 (-2)
B2AV800LSK-1(-2)
Note:
The model number suffixes of -1 and -2 denote 110VAC and 220VAC ratings
respectively. Model number without the suffix hereafter represents both ratings.
The main PCB (ASM0021) has a DIP switch (S1), which needs to be set correctly to
differentiate models. Only trained personnel should change these settings.
If jumper 3-4 of 2x3 jumper header (JP4) is open, then the table is equipped with
DC24V/14Ah lead acid batteries. If it is closed, then it is equipped with DC24V/17.4 Ah
lithium iron phosphate batteries.
Remarque:
Les suffixes -1 et -2 du numéro de modèle indiquent respectivement une alimentation de
110VCA et de 220VCA. Tout numéro de modèle indiqué sans suffixe représente les
deux types d'alimentation.
La carte de sacs imprimés principale (ASM0021) comprend un commutateur DIP (S1)
qui doit être réglépour différencier les modèles. Seul le personnel qualifié devrait
modifier ces paramètres.
Si le cavalier 3-4 du système 2x3 (JP4) est ouvert, la table est équipée de batteries au
plomb-acide de 24VCC/14Ah. Si ce cavalier est fermé, la table est équipée de batteries
au lithium phosphate de fer de 24VCC/17,4Ah.
Tabletop
Kidney
Sliding
Elevator
N/A
N/A
N/A
N/A
N/A
N/A
N/A
N/A
3-1
DIP Switch (ASM0021)
Floor
Locks
No. 3
manual
manual
On
manual
manual
auto
auto
On
auto
auto
B2AV800 Series Surgical Table
IM-B2AV800-E01
No. 4
On
Off