Lock Table In Place - Nuvo B2AV800 Series Instruction Manual

Surgical table
Table of Contents

Advertisement

1. Push Hand Pendant to firmly clip it on
an appropriate side rail. Do not only
hang it on the side rail.
Note:
When removing the Hand Pendant
from the side rail, pull it straight out.
Do not twist it from side to side, or the
metal clip could be damaged.
Remarque:
Lorsque vous retirez le boîtier de commande manuelle du rail latéral, tirez-le vers vous. Ne
le tournez pas de gauche à droite, car cela pourrait endommager l'attache en métal
2. Locate the receptacle on column top
underneath the seat section.
3. Insert the connector all the way into the
receptacle, and then tighten its screw
to prevent the connection from
becoming loose.
4.3

Lock Table in Place

SHOCK HAZARD: Monitor the electrically conductive properties of the floor locks (all
models) and casters (B2AV800S(K) and B2AV800(K)). Measure their resistance level
according to the requirements of NFPA 99 before placing table in use. Subsequent
routine testing should be performed at least once every three months.
RISQUE DE DÉCHARGE ÉLECTRIQUE: Surveillez les propriétés de conductivité
électrique des dispositifs de verrouillage au sol. Mesurez leur niveau de résistance
selon les exigences de la norme NFPA 99 avant de commencer à utiliser la table. Des
tests de routine doivent être effectués au moins une fois tous les trois mois.
Figure 4-1:
Figure 4-2:
WARNING! AVERTISSEMENT!
4-3
B2AV800 Series Surgical Table
IM-B2AV800-E01