Every model of the B2AV800 Series along with appropriate accessories provides excellent patient
positioning for both traditional surgery and modern endoscopic surgery. Constructed from
aluminum alloy, stainless steel, and other high strength steel, the B2AV800 Series with its
powerful electro-hydraulic system is rigid enough to position patients up to 704 lbs (320 kg) in
normal and reverse orientation. Ask your NUVO sales representative for available accessories
sold separately from the table.

3.2 Product Classification and Identification

B2AV800 Series surgical table's electrical
insulation structure is categorized as CLASS
1 TYPE B with IPX4 fluid ingress protection.
Each individual surgical table has a unique
serial number printed on the identification
plate, which is located on the leg side of the
column as illustrated to the right.
3.3

Requirements

3.3.1 Operating Environment Requirements

 Temperature: 0°C ~ 40°C (32°F~104°F)
 Relative humidity: 20 ~ 80 %RH
 Atmosphere Pressure: 700 ~ 1060 kPa

3.3.2 Power Requirements

 AC 110V / 5A, 300W, 50 / 60Hz or
 AC 220V / 3A, 300W, 50 / 60Hz; single phase
Note:
1. Plug the power cord into an equivalent electrical outlet marked "Hospital only" or "Hospital
Grade" for effective electrical grounding.
2. B2AV800 Series surgical tables are shipped with standard electrical specifications and
configurations. If a change needs to be made to the electrical system, contact NUVO for
instructions on the correct procedure.
Remarque:
1. Branchez le cordon d'alimentation dans une prise électrique équivalente marquée « Hôpital
uniquement » ou « Qualité hôpital » pour obtenir une mise à la terre efficace
Figure 3-1: Identification Plate Location
3-2
B2AV800 Series Surgical Table
IM-B2AV800-E01