Display/Code
13 When rotary valve keeps turning but
receives no signal from U1~U4 of
ASM0027
Rotary valve position number not
14
matching position code
Rotary valve position code correct but
15
stops too early, and fails to position
after programmed compensation.
Rotary valve position code correct but
16
stops too late, and fails to position
after programmed compensation.
Not used
17
Not used
18
Not used
19
Not used
20
Batteries are not being charged (i.t.)*
21
Not used
22
Using LOCK or UNLOCK button of
23
B2AV800L(S) hand pendant on
B2AV800(S)(K) table
Using Slide button of hand pendant on
24
Non-slide table
25 Using T1000 Series hand pendant on
B2AV800 Series tables.
Not used
26
Voltage is too high (i.t.)*
27
Symptom
5-13
B2AV800 Series Surgical Table
IM-B2AV800-E01
Cause
1. Short circuit within the angle
system.
2. Rotary valve motor damaged
3. ASM0023 driver PCB
damaged
4. ASM0027 or its cable
damaged
ASM0027 or its cable damaged
Timing parameter is set too low
Timing parameter is set too high
1. Loose battery connector
2. Battery reached life span or
over discharged
3. Broken PCB (ASM0021)
1.
Wrong hand pendant
2.
Wrong DIP setting
1.
Wrong hand pendant
2.
Wrong DIP setting
Wrong hand pendant.
Wrong setting