2.
Release the button when the desired position is reached.
3.
For LOCK or UNLOCK, press the button and release it after the function has been
activated. The table will automatically LOCK or UNLOCK (for model w/ L suffix).
4.
When the table begins to move on its own while using the hand pendant, press the
OFF button right away and call for service.
5.2.1.1

Level Button Operation

Press Level button (#27, Figure 5-2) until tabletop has reached level, then let go of the
button.

5.2.1.2 Normal Orientation Operation

Press Normal button (#4, Figure 5-2) and its green LED will be lit. All buttons will
automatically switch to normal orientation.

5.2.1.3 Reverse Orientation Operation

Press Reverse button (#6, Figure 5-2) and its green LED will be lit. All buttons will
automatically switch to reverse orientation.

5.2.1.4 Battery Level Indicator

Besides indicating level of battery power, it also indicates the need to change the
battery or batteries when all three LEDs within the battery symbol are flashing.

5.2.1.5 Interlock LED

As a protective feature, when the Interlock indicator (#17, Figure 5-2) is lit (buzzer
beeping at the same time), it means one of movements (table down, Trendelenburg,
Reverse Trendelenburg, tilt, leg section down, back section down or slide out) has
stopped because it is colliding with something, and the button of the above said
movements is disabled. It could also mean that one of the above said movements
has reached its maximum distance or angle.

5.2.1.6 Service LED

While operating Hand Pendant, if Service indicator (#18, Figure 5-2) is lit (buzzer
beeping at the same time), it could mean override system is on. Turn it off and the
Service LED should disappear. Other than that, when Service LED is on while using
the hand pendant, then a problem has occurred that renders certain component(s)
unusable. At this time, the 10-segment LED Battery Indicator (#8, Figure 5-3) could
display Service Code. For other events, which the Service LED is lit only while a
button is pressed; press the button while looking for the Service Code display. Let go
of the button to exit from the Service Code display. Inform your finding to the NUVO