Nuvo B2AV800 Series Instruction Manual page 15

Surgical table
Table of Contents

Advertisement

CAUTION:
interference or other types of interference.
Before using high-frequency surgical equipments, cardiac defibrillators or cardiac
defibrillator-monitors, please refer to their instruction manual.
Use only NUVO's mattress pads in order to avoid risk of electrostatic shock.
When the power cord is damaged in any way, replace it immediately or else use the
table's internal electrical power source until it can be replaced.
The antistatic pathway is: Mattress Pad→Tabletop→Column→Base→Floor Lock
Pads→Floor. Therefore, the anti-static properties of the table depend upon the use of
the recommended NUVO mattress pads.
If the surgical drape hangs over the articulation points or the articulation covers (stainless
steel plates covering the joints), movement could cause damage to the table. Tuck the
drapes between the mattress pad and tabletop in order to prevent this.
Static electricity could damage the IC. Avoid any bodily contact with the electronic
systems. Use probes if necessary. Personnel must use a grounding device when
probing circuits or connectors, or else immediate electrical failure could result.
There is only one way to insert the connector of the hand pendant into its receptacle on
the table. The connector must be inserted straight into the receptacle without twisting it,
or else pins inside the receptacle will be bent and rendered unusable.
Since collision with the table joint covers may damage them, keep stretchers or other
heavy objects away from the covers.
Read the entire instructional manual before operating the table.
MISE EN GARDE :
 Seuls les nettoyants et désinfectants recommandés doivent être utilisés sur les tables
NUVO. N'utilisez pas d'alcool pour procéder au nettoyage de la table, car ses propriétés
de nettoyage et de désinfection sont insuffisantes et n'utilisez pas de composés
phénoliques qui risquent de brûler la peau du patient s'ils sont mal rincés.
 Acheminez le cordon du boîtier de commande manuelle suspendu et le cordon de la
pédale de commande optionnelle de manière à éviter tout point de pincement susceptible
de les endommager.
 Pour éviter d'endommager le boîtier de commande manuelle suspendu, attachez-le au rail
latéral de la table lorsqu'il n'est pas utilisé.
 Veuillez lire attentivement l'étiquette du flacon aérosol et suivez toutes les instructions et
1-7
B2AV800 Series Surgical Table
IM-B2AV800-E01

Advertisement

Table of Contents
loading

Table of Contents