Nuvo B2AV800 Series Instruction Manual page 77

Surgical table
Table of Contents

Advertisement

2.
Remove fuse socket caps (#1 and #2, Figure 5-3) with a screwdriver. The cap is not
screwed in, therefore do not try to unscrew it or else the cap may become damaged.
Just push it in the cap and turn slightly to unlock it and use hand to pull it out.
Figure 6-2
3.
Replace blown fuse(s) with new one(s) according to the specifications below.
NUVO Item Id
EF0002
EF0003
4. Push and turn slightly to lock the fuse socket caps.
Specifications
6*32 3A/250V (time-lag)
6*32 5A/250V (time-lag)
6-9
B2AV800 Series Surgical Table
IM-B2AV800-E01
AC Voltage
220V~240V
110V~120V