Using The Override System For Table Positioning - Nuvo B2AV800 Series Instruction Manual

Surgical table
Table of Contents

Advertisement

Display/Code
28
29
30
31
(it)*: Initiating test of system, which happens when the system is being power up after
it has been reset by being disconnected from all power sources.
5.2.2
Using Override System for Table Positioning
PINCHING HAZARD: Extreme tabletop articulation creates pinch points. Before
operating the table, review illustrations in Figure 5-1, PINCH POINTS.
PINCHING AND TIPPING HAZARD: If user is not completely familiar with the controls
for patient positioning and table operation, patient could sustain injury.
EXPLOSION HAZARD: Do not use surgical table with flammable anesthetics.
UNEXPECTED MOVEMENT HAZARD: If the surgical table begins to move on its own
when the override control toggle switch is put in either the up or down position,
immediately put the switch back in the neutral position, discontinue use of the table,
and call a technician for service.
UNEXPECTED TABLE MOVEMENT HAZARD: When using any control, please
operate one button or switch at a time to avoid unexpected table movement.
RISQUE DE PINCEMENT: L'articulation excessive du dessus de table crée des points
de pincement. Consultez les illustrations des figures 5-1, « POINTS DE
PINCEMENT » avant l'utilisation la
RISQUE DE PINCEMENT ET DE BASCULEMENT: Si l'utiliisateur n'a pas une parfaite
connaissance des contrôles de positionnement du patient, le patient subir des
blessures.
Symptom
Voltage is too low (i.t.)*
Not used
Not used
Wrong DIP Switch setting (i.t.)*
WARNING!
AVERTISSEMENT!
table.
5-14
B2AV800 Series Surgical Table
IM-B2AV800-E01
Cause
Wrong setting