GENERATOR EXPOSURE
TABLES
Table 1 following shows mAs values resulting from preselected mA and ms values (3 point
operation). This table also shows the range and interrelation of these loading factors. For
example, if in AEC operation it is known that the preset backup mAs limit is 50 mAs and 100
mA is chosen for that technique, then the maximum length of exposure possible (at which time
the backup mAs limit is reached) is 500 ms.
In the above example, read down the 100 mA column to the 50 mAs value. Then read
across to the left side of the page to the 500 ms value, this will be the maximum time allowed
for the stated conditions.
Please review the note at the bottom of page 50. This details AEC backup options that
are available.
IEC 60601-2-7 requires that this information be provided in a form suitable to be displayed at a
prominent location on or near the control console of the generator. To this end, an extra copy
of table 1 is provided on pages 46 and 47 which may be photocopied as required, or removed
from this manual and placed in a suitable location as per local requirements.
Indico 100 Series Operator Manual
740953-00
Rev. A
7