Use these steps to edit the programmed APR techniques.
NOTE: BEFORE MAKING CHANGES TO THE DEFAULT APR TECHNIQUES, PLEASE
RECORD ALL CURRENT DEFAULTS. THESE DEFAULTS MAY BE RECORDED IN
COPIES OF THE APR PROGRAMMING WORKSHEET AT THE END OF THIS CHAPTER.
1.
From the GENERATOR SETUP menu select APR EDITOR.
2.
Press the APR EDITOR button to toggle between ENABLED and DISABLED.
DISABLED does not allow editing and then saving the changed APR technique(s).
ENABLED allows the operator to change the default APR techniques, and then save the
changes. The APR will then default to the newly saved techniques.
The generator stores the last APR editor setting prior to being switched OFF. If
the APR editor was previously ENABLED, APR changes may subsequently be
made and then saved in normal operating mode without the need to manually set
the APR editor to ENABLED. To disable this feature, the APR editor must be set to
DISABLED.
3.
With the APR EDITOR set to ENABLED, return to normal operation by pressing EXIT
SETUP.
4.
The following menu is displayed. This is a typical menu, actual anatomical views may
have been reprogrammed, and the screen may differ slightly depending on generator
model.
Indico 100 Series Operator Manual
* GENERATOR SETUP *
UTILITY
EXIT SETUP
TUBE 1
0%HU
SKULL
THORAX
ABDOMEN
PELVIS
160MS
740953-00
APR EDITOR: DISABLED
* APR TABLES *
LOWER EXTREMITY
UPPER EXTREMITY
G.I.
SPINE
Rev. A
Programming the Console
1:16
Page 33