CPI INDICO 100 Series Operator's Manual page 10

Hide thumbs Also See for INDICO 100 Series:

Advertisement

2
Safety and Specifications
OUTPUT PARAMETERS
kVp range:
kVp accuracy:
mA range:
mA accuracy:
mAs range:
Time range:
Lowest current - time
product:
Loading factors which
will allow lowest
current - time product:
Page 10
Radiography 40 to 150 kVp.
Fluoroscopy 40 to 125 kVp.
± (5% + 1) kVp.
Radiography 10 to 320 mA (30 kW), 10 to 400 mA (37.5 kW), 10
to 630 mA (50 kW), 10 to 800 mA (65 kW), and 10 to 1000 mA (80
kW).
Fluoroscopy 0.5 to 6.0 mA.
± (5% +1.0) mA for all mA values.
Tube dependent, max 1000 mAs.
Radiography 2.0 to 6300 ms.
Fluoroscopy 0 to 5 or 0 to 10 minutes.
0.5 mAs.
All mA values ≤ 250 mA. Refer to table 1 in section 7.
Rev. A
Indico 100 Series Operator Manual
CPI Canada Inc
740953-00