Gilson logo engraved on their collar, ensuring that you
have a genuine Gilson product. Plastic tips are for a
single application – they must not be cleaned for reuse.
PIPetman G can also be used with the main tip brands.
Fitting the tips
To fit a new PIPETMAN DIAMOND Tip, push the tip-holder
into the tip using a slight twisting motion to ensure a
firm, airtight seal.
For the P2G and P10G models equipped with stainless
steel tip-ejector,
A dual-position adapter (plastic) is
required to fit DL10 tips (long tips) or
D10 tips (short tips).
P2G and P10G models are delivered
with the adapter in place, ready to
use DL10 tips. If D10 tips are used, the
adapter must be repositioned in the
shorter slot as follows:
Pull the adapter down from the
metallic rod.
Turn the adapter through 180°C.
Refit the adapter so that the end
of the metallic rod engages the
shorter slot of the adapter.
For the P2G and P10G models equipped with plastic
tip-ejector,
A tip-ejector extension is supplied to fit with
D10 tips (short tips).
To fit a tip-ejector extension:
Slide the extension over the tip-holder.
Push the extension firmly onto the end
of the tip-ejector until it clicks into place.
To remove a tip-ejector extension:
Gently twist the extension.
Pull it away from the pipette.
Both dual-position adapter and tip-ejector
extension are autoclavable .
Short tips Long tips