0.5 μL to 1200 μL, including four V-ring (VR) models, covering a volume range from 20 μL to 300 μL. PIPETMAN L V-ring (VR) multichannel models are equipped with V-rings at the bottom of the tip holders. They create an...
Page 3
Description Single Channel Models Multichannel Models Color-coded push button Tip ejector Thumbwheel Ejector support Tip ejector button Cover Handle PIPETMAN® DIAMOND Tip(s) Connecting nut V-rings (for VR multichannel models only) Ejector clip Graduated cover for fixed Tip holder(s) volume models only Figure 2 PIPETMAN®...
Specifications PIPETMAN L is a high-quality pipette that offers excellent Maximum accuracy and precision. The figures given in the Gilson Permissible Errors tables were obtained using PIPETMAN® DIAMOND Tips. These values are guaranteed only when genuine PIPETMAN DIAMOND Tips are used.
Page 5
PIPETMAN® L SINGLE CHANNEL—VARIABLE VOLUME MODELS Maximum Permissible Errors Vol. Gilson ISO 8655 (µL) Systematic Random Systematic Random Systematic Random Random Systematic Error Error Error Error Error Error Error Error (µL) (µL) (%CV*) (µL) (µL) (%CV*) P2L (P/N FA10001M AND FA10001P) WITH D10 AND DL10 TIPS ±...
User Adjustments (Fixed Volume Models Only) PIPETMAN L fixed volume models are factory calibrated using distilled water and very high precision balances. The nominal volume may be adjusted slightly to compensate for liquids of different density or viscosity.
D5000 and D10mL PIPETMAN DIAMOND Tips do not have any filter. P5000L and P10mLL models are sold with NOTICE a bag of 10 filters. For more details, refer to PIPETMAN L User’s Guide LT801575. Multichannel Models PIPETMAN DIAMOND Tips: TIPACK AND TOWERPACK PIPETMAN DIAMOND Tips are best fitted with the ROCKY RACK™...
Page 9
Fitting the Tip Ejector Adapter and the Tip Ejector Extension Single Channel Models For P2L and P10L pipettes equipped with a stainless steel tip ejector, a dual-position adapter is required to fit and eject DL10 tips (long collar tips) and D10 tips (short collar tips).
Wait one second (time depends on model, for more details, please refer to PIPETMAN L User’s Guide LT801575), and then withdraw the pipette tip from the liquid. You may wipe any droplets away from the outside of the tip using a medical wipe;...
Replacing spare parts ● Greasing the piston assembly Cleaning and Decontamination PIPETMAN L is designed so that the parts normally in contact with liquid contaminants can easily be cleaned and decontaminated, and the variable volume models are autoclavable without disassembly.