Service and troubleshooting
May 2021
7.3

Troubleshooting incorrect level readings

7.3.1
Reported level is too high or low
Figure 7-1: Symptom
A. Level
B. Time
C. Actual level
D. Reported level
Possible cause
Incorrect tank geometry configuration.
Recommended actions
Related information
Reference height
Analyze the echo peaks
Adjust the general threshold
Restore to factory settings
48
A
C
D
Verify the tank geometry parameters are configured correctly (especially the
Reference Height).
Analyze the Echo Peaks and check General Threshold.
Restore default settings and reconfigure the device.