Emerson Rosemount 1408A Reference Manual page 49

Level and flow transmitter
Hide thumbs Also See for Rosemount 1408A:
Table of Contents

Advertisement

Reference Manual
00809-0200-4480
7.3.2
Level is stuck in measuring range
Figure 7-2: Symptom
A. Level
B. Time
C. Actual level
D. Reported level
Possible cause
Disturbing object in the tank.
Recommended actions
Related information
Analyze the echo peaks
Adjust the general threshold
Mounting position
Reference Manual
A
C
D
Analyze the Echo Peaks and check General Threshold.
Remove the disturbing object.
Put an inclined metal plate on top of the disturbing object.
Move the transmitter to another position.
Service and troubleshooting
B
May 2021
49

Advertisement

Table of Contents
loading

Table of Contents