Troubleshooting by user

Failure message
Problem with centrifuge
Chamber overtemp
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Incorrect bucket ID
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Unapproved rotor
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Rotor ID failure
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Lid blocked
Centrifuge does not open
Motor overtemp
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
PCB overtemp
The centrifuge cannot be operated.
The run does not start or the centrifuge runs
down without being braked.
Thermo Scientific
If problems occur other than those listed in this table, the authorized customer service
representative must be contacted.
Possible causes and cures
Overheating in chamber.
Check the function of the refrigeration unit.
Clean the air inlet for the condenser.
Restart the centrifuge.
If an error message appears again, inform your service
technician.
Bucket code is undefined for the rotor detected, check the set
points for the given bucket code.
Is it permitted to use the current bucket in the rotor currently
mounted?
Restart the centrifuge.
If an error message appears again, inform your service
technician.
Rotor code is not in the rotor table.
Is it permitted to use the rotor currently mounted in this device?
Restart the centrifuge.
If an error message appears again, inform your service
technician.
The rotor could not be identified.
Check to see if the rotor is properly installed.
Restart the centrifuge.
If an error message appears again, inform your service
technician.
Restart the centrifuge.
The emergency lid release enables you to retrieve your samples.
If an error message appears again, inform your service
technician.
Restart the centrifuge.
If an error message appears again, inform your service
technician.
Restart the centrifuge.
If an error message appears again, inform your service
technician.
7
Troubleshooting
Troubleshooting by user
Sorvall Legend X1R
7-3