On the back of the instrument there is a fuse board with 4 fuses (→ "Fig. 59"):
1. Unscrew the sleeve of the defective fuse using the screwdriver.
2. Remove the sleeve and fuse.
3. Insert the defined fuse into the sleeve, then use the screwdriver to screw the sleeve back in as
far as it will go.
Fig. 59
Fuse
Function/protection
F1
Empty
F2
Coarse feed
F3
Processor board power supply
F4
Heaters
F5
Peltier element
9.3.3 Replacement of the UVC lamp
Warning
Switch the instrument off and disconnect the power plug before replacing the UVC lamp! Wear suitable
safety gloves and safety glasses.
If the lamp is broken, it must be replaced by the service technician, as the replacement involves a high
risk of injury.
Leica CM1860/CM1860 UV
Type
T 1.6 A
T 1.0 A
T 6.25 A
T 4 A
A UVC lamp has an estimated lifetime of approx. 9000 hours.
Each on/off switching cycle reduces the lamp life by approx. one hour plus burning
time (30 minutes or 180 minutes respectively).