Temperature control button

On the back of the instrument there is a temperature control button (→ Fig. 54-1). If the temperature of
the cryochamber exceeds 60 °C the switch responds automatically and turns the instrument off.
1
Fig. 54
Possible causes and remedies:
1. Temperature of the immediate surroundings is constantly higher than 40 °C.
 Drop the temperature of the immediate surroundings.
2. When setting up the instrument, the minimum clearance
requirements)
 Observe the minimum clearance.
3. The air inlet openings of the condenser are dirty.
 Clean the air inlet openings (→ Fig. 57-5).
After eliminating the possible sources of error, push the temperature control button
the instrument back on and make it ready to operate. If the instrument does not respond, contact service.
66
to walls and furniture was not kept.
(→ P. 22 – 4.1 Installation site
(→ Fig. 54-1)