Mobile Heat Extractor - Application - Leica CM1860 Instructions For Use Manual

Hide thumbs Also See for CM1860:
Table of Contents

Advertisement

10
Ordering Information, Components and Consumables
Fig. 116
Fig. 117

10.2 Mobile heat extractor - application

Specimen freezing with the freezing shelf can be accelerated by the additional use of a heat extractor.
1. Store the heat extractor in the cryochamber.
2. Hold it on the specimen surface to accelerate freezing.
3. Take it away once the specimen is entirely frozen.
Note
Recommendation:
It is recommended to precool the heat extractor in liquid nitrogen or other refrigerant.
Warning
Caution! There is a risk of injury due to freezer burn!
102
UVC fluorescent tube 4-pin
Order No.
Storage system, movable
for installation in the front part of the cryostat for
cooled storage of preparation aids.
Order No.
14 0471 57384
14 0491 46750
Version 2.0, Revision N