Instrument Controls; Control Panel Fields On The Cm1950 - Control Panel Field 1 - Leica CM1950 Operating Manual

Uvc disinfection effective for destroying clinically relevant pathogens
Hide thumbs Also See for CM1950:
Table of Contents

Advertisement

7.

Instrument Controls

7.1
Control panel fields on the CM1950 – Control panel field 1
Display of actual and target chamber
temperatures
Fig. 32
Button to
Button to
enable/disable
enable/disable
extraction
the Peltier
system,
element
Extraction
force intensity
selectable
from 1-5 (see
following page
for more
information)
30
Display of actual and target specimen
head temperatures
+/- buttons to
select
chamber
temperature
Max-Cool button to
select the maximum
low temperature of
the specimen head
directly (-50 °C)
EMERGENCY STOP switch to the right of control panel field 1 (motorized
instruments only)
For danger situations during motorized sectioning.
• Immediate stop of the sectioning process–motor stops–LED in E-Stop is il-
luminated in red.
• Turning in direction of arrow cancels the stop – LED in E-Stop goes out.
• Select single stroke (Single) or continuous stroke (Cont.) operating mode
again.
+/- buttons to
+/- buttons to
select
select real
specimen
time
head
temperature
+/- buttons to
select defrost
Operating Manual, Rev C, 1v3 – 04/2010
Display of real time, defrost time
and error messages.
"Melting
Snowflake"
button to
activate manual
defrost
time
ON/OFF button of
illumination
Press the key button to
lock/unlock the entire
keypad.
(Refer to following page
for activation of the
specimen head)
UVC
disinfection,
(short time 30 min,
long time 180 min)