AMU Rescon
Program List and Explanations
4.3 Logger
4.3.1
Interval
1 s
Time
25 min
4.3.2

5 Installation

5.1 Sensors
5.1.1
5.1.2
50
The instrument is equipped with an internal logger. The logger data
can be downloaded to a PC using the built-in RS232 interface.
The logger can save approx. 1500 data records. The Records con-
sists of: Date, time, alarms, measuring value, raw value (MΩ), case
temperature, flow.
Log Interval: Select a convenient log interval. Consult the table be-
low to estimate the max logging time. When the logging buffer is
full, the oldest data record is erased to make room for the newest
one (circular buffer).
Range: 1 Second to 1 hour
5 s
1 min
2 h
25 h
Clear Logger: If confirmed with yes, the complete logger data is de-
leted. A new data series is started.
Flow: Select the type of flow cell. Possible flow cells:
Mode
None
Q-flow
Q-Hflow
Selection of the suitable flow cell.
Flow cell type
B-Flow
Q-Flow or QV-Flow
Q-Hflow or QV-Hflow
Meas. Mode: The two measuring modes Conductivity or Resistivity
are available.
5 min
10 min
5 d
10 d
B-Flow
Q-Flow
Mode
None
Q-flow
Q-Hflow
30 min
1 h
31 d
62 d
QV-flow