Fine Adjust - Swann AMU Rescon Operator's Manual

Table of Contents

Advertisement

AMU Rescon
Maintenance
Start transmit-
ter check
6.5.
A-96.250.351 / 230320
NOTICE: If Transmitter Check is switched on, the main menus
<Operation> and <Installation> are not accessible.
Maintenance
Simulation
Set Time07.05.12 11:06:17
Transmitter checkoff

Fine adjust

Maintenance
Simulation
Transmitter check
Set Time07.05.12 11:06:17
off
Transmitter Checkoff
on
Fine adjust
Maintenance
Simulation
Set Time07.05.12 11:06:17
Transmitter checkon
Fine adjust
Fine Adjust
The function fine adjust is only available if the AMU Rescon is set
to the measuring mode resistivity.
The function "Fine Adjust" is performed automatically every night at
00:30 h.
It is also possible to start the function "Fine Adjust" manually in the
menu <Maintenance/Fine adjust>.
1 Navigate to menu <Maintenance> /
3.3
<Transmitter check>.
2 Press [Enter].
3 Select <on> and confirm with [En-
3.3
ter]
 The Transmitter check is now
3.3
active
4 Abandon the menu by pressing the
exit key twice.
5 Compare the displayed resistivity
or conductivity value and the tem-
perature value with the value print-
ed on the test resistor.
37