Connecting Sample Inlet and Outlet

Sample Inlet

For the sample inlet a stainless steel tube is connected to the Swa-
gelok fitting at the flow cell. Cut the tube to length and deburr it. The
tube must be straight and free from blemishes for approximately
1,5 x tube diameter from the end.
Lubrication with lubricating oil, MoS2, Teflon etc. is recommended
for the assembly and reassembly of bigger sized unions (thread,
compression cone).
1 Insert the compression ferrule [C] and the compression
cone [D] into the union nut [B].
2 Screw on the union nut onto the body, do not tighten it.
3 Push the stainless steel pipe through the union nut as far as it
reaches the stop of the body.
4 Mark the union nut at 6 o'clock position.
5 While holding the fitting body steady, tighten the nut union 1¼
rotation using an open ended spanner.
A
6

Sample Outlet

FEP flexible tube 6 mm. Connect the tube to the Serto elbow union
and Insert it into an atmospheric drain of sufficient capacity.
Max. tube length is 1.5 m. Do not connect longer tubes.
B
C D
E
F
A
Tube
B
Union nut
C
Compression
ferrule
D
Compression
cone
E
Body
F
Tightened
connection
A-96.250.351 / 230320