3.2 Set Time
3.3 Transmitter check
3.5 Fine adjust

4 Operation

4.1 Sensors
4.1.1
4.1.2
4.2 Relay Contacts
A-96.250.351 / 230320
Alarm Relay:
Relay 1:
Relay 2
Signal Output 1:
Signal Output 2
At the absence of any key activities, the instrument will switch back
to normal mode after 20 min. If you quit the menu, all simulated val-
ues will be reset and the transmitter reboots.
Adjust date and time.
Transmitter check is used to check the accuracy of the AMU trans-
mitter's electronic parts. The test fulfills the requirements of the
United States Pharmacopeia (USP). See
QC-Kit Test Plug, p.
The function fine adjust is only available if the AMU Rescon is set
to the measuring mode resistivity.
The function "fine adjust" performs an internal resistor adjustment.
Filter Time Constant: Used to damp noisy signals. The higher the
filter time constant, the slower the system reacts to changes of the
measured value.
Range: 5–300 Sec
Hold after Cal: Delay permitting the instrument to stabilize again af-
ter calibration. During calibration plus hold-time, the signal outputs
are frozen (held on last valid value), alarm values, limits are not ac-
tive.
Range: 0–6'000 Sec
See
Relay 1 and 2, p.
Active or inactive
Active or inactive
Active or inactive
Actual current in mA
Actual current in mA
35.
22.
Conductivity-Resistivity
49