Maintenance of Sensor

Clean Sensor

The Swansensor RC-U is largely maintenance free. However, de-
pending on the application, it can be contaminated, which may
cause problems.
The Swansensor RC-U is available in the following 2 different ver-
sions:
 Sensor with fixed cable
 Sensor with plug
NOTICE: Sensor with fixed cable
• To avoid damage of the sensor cable due to torsion when
screwing the sensor out of the flow cell, disconnect the cable
from the terminals of the AMU transmitter.
1 Open the transmitter housing.
2 Disconnect the sensor cable from the terminals.
3 Remove the sensor cable from the transmitter housing.
4 Unscrew and remove the sensor [A] from the flow cell block [B],
use a monkey spanner.
5 Remove the teflon tape from the sensor thread.
6 Clean the sensor with soap water.
7 Rinse the sensor well with high purity water.