AMU Rescon
Instrument Setup
4.
4.1.
4.2.
Flow
Measurement
Meas. Mode
A-96.250.351 / 230320
Instrument Setup
Establish Sample Flow
1 Open the flow regulating valve.
2 Wait until the flow cell is completely filled.
3 Switch on power.
NOTICE: To allow a precise measurement in the range
between 18 –18.18 MΩ, set the sample flow to 70–100 l/h.
Programming
Set all necessary parameters in menu 5 <Installation>, further infor-
mation about sensor parameters see
Flow measurement
Measuring Mode
USP Operating Mode
Sensor parameters
Temperature compensation
Program all parameters for external devices (interface, re-
corders, etc.). Program all parameters for instrument opera-
tion (limits, alarms). See
explanations, see
Menu 5.1.1
Select the flow sensor you are using:
None
Q-Flow
Q-HFlow
Q-Flow can be used for flow rates up to 25 l/h.
Q-HFlow can be used for flow rates up to 120 l/h.
Menu 5.1.2
Select Resistivity or Conductivity according to your application.
5.1 Sensors, p.
Program Overview, p.
Program List and Explanations, p.
50.
42, for
47.