AMU Rescon
Installation
3.6.
Installation
requirements
3.7.
3.8.
3.9.
A-96.250.351 / 230320
Power supply
CAUTION
Do not apply power to the transmitter until all electrical connec-
tions have been made.
A
B
C
The installation must meet the following requirements:
Mains cable according to standards IEC 60227 or IEC 60245;
flammability rating FV1
Mains equipped with an external switch or circuit-breaker
– near the instrument
– easily accessible to the operator
– marked as interrupter for AMU Rescon
Sensor
Connect the sensor to the AMU transmitter according to the con-
nection diagram, see
For sensor settings, see
Flow Meter
Connect the flow meter (if any) to the AMU transmitter according to
the connection diagram, see
Input
NOTICE: Use only potential-free (dry) contacts.
Terminals 10/11
For programming see
A
Power supply connector
B
Phase conductor
C
Neutral conductor
Electrical Connection, p.
Programming, p.
Electrical Connection, p.
Program List and Explanations, p.
19.
25.
19.
47.