3.10. Relay Contacts

3.10.1 Alarm Relay

3.10.2 Relay 1 and 2

3.11. Signal Output 1 and 2 (current outputs)

22
NOTICE: Max. load 100 mA/50 V
Alarm output for system errors. Error codes see
Terminals
a)
NC
5/6
Normally
Closed
a)
NO
5/6
Normally
Open
a) As defined when ordering
NOTICE: Max. load 100 mA/50 V
Relay 1: Terminals 1/2
Relay 2: Terminals 3/4
For programming see
Installation
NOTICE: Max. burden 510 Ω
If signals are sent to two different receivers, use signal isolator
(loop isolator).
Signal output 1: Terminals 13 (+) and 12 (-)
Signal output 2: Terminals 14 (+) and 12 (-)
For programming see
Installation
Description
Active (opened) during normal
operation.
Inactive (closed) on error and loss of
power.
Active (closed) during normal
operation.
Inactive (opened) on error and loss of
power.
Program List and Explanations, p.
Program List and Explanations, p.
Error List, p.
39.
47, Menu
47, Menu
A-96.250.351 / 230320