Conductivity-Resistivity Qc-Kit Test Plug; Introduction - Swann AMU Rescon Operator's Manual

Table of Contents

Advertisement

AMU Rescon
Maintenance
6.4.
6.4.1
Test mode
A-96.250.351 / 230320

Conductivity-Resistivity QC-Kit Test Plug

The QC-Kit test plug fulfills the requirements of the United States
Pharmacopeia (USP).

Introduction

The test plug consists of 2 high precision resistors for conductivity
and temperature.
The plug has a resistor of
 5000  (+/- 1%) for temperature
 181800  (+/- 1%) for conductivity.
NOTICE: Keep test kit absolutely dry!
Several variables enter into the calculation of specific conductivity.
Values like the cell constant, temperature compensation algorithm,
cable length and temperature correction of the sensor are stored in
the memory of the conductivity instrument.
For a quick and easy test, the AMU Rescon transmitter has the test
mode <Transmitter check> that enables testing without changing
any of these values. When leaving the test mode, all parameters
will be set back to their original value.
35