Download Print this page

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1

Quick Links

IMPORTANT! Do not provide these setup
instructions to the patient. Only qualified
medical professionals should access these
device settings.
Follow these instructions to access the device
settings and prepare the device for patient use.
All settings have pre-set values as shown below.
First -
With the device pow-
ered and turned off, press and
hold the
DOWN
and
DELAY
buttons.
Second -
While holding the
two buttons, press the
ON/
OFF
button. The display will
show "Clinical Menu". You
can now access the Clinical
Setup Menu.
Menu Item
Description
AutoAdjust Mode Clinical Menu
Language
Select device language.
Selection
Operating
Adjust mode to AutoAdjust or CPAP as prescribed by
Mode
doctor.
Range: AutoAdjust, CPAP
Upper
Maximum pressure level that will be administered by the
Limit
device.
Range: Lower Limit to 20 cmH
Lower
Minimum pressure level that will be administered by the
Limit
device.
Range: 3.0 cmH
0 to Upper Limit, increments of 0.5
2
Delay
During delay the pressure remains at Lower Limit and
Time
the algorithm continues to detect events, but no pres-
sure changes are made.
Range: 0-45 minutes, increments of 5
Pressure
The device operating pressure may be displayed in one
Unit
of three pressure units: cmH
Tubing
The device is supplied with air supply tubing of 1.8 m
Length
(6 ft) in length. The device may also be operated with air
supply tubing of 3 m (10 ft). This setting should be
updated to match the tubing length for maximum
pressure accuracy.
Compliance
Cumulative hours of "while breathing" time.
Meter
Advanced
Press the  button to enter. Reference the Advanced
Menu
Submenu section for details.
Enable
Press the  button to enter. Reference the Enable
Menu
Submenu section for details.
Lockout
Press the  button to enter. Reference the Lockout
Menu
Submenu section for details.
CPAP Mode Clinical Menu
Language
Select device language.
Selection
Operating
Adjust mode to AutoAdjust or CPAP as prescribed by
Mode
doctor.
Range: AutoAdjust, CPAP
Set
Set to CPAP pressure prescribed by doctor.
Pressure
Range: 3-20 cmH
O
2
Delay
Select period of time to stay at delay pressure. This
Time
delay is patient activated.
Range: 0-45 minutes, increments of 5
Delay
If prescribed, set to a level that is comfortable for the
Pressure
patient to fall asleep. This level of pressure is maintained
for the first portion of the delay time and rises to the set
pressure in the last 10 minutes of the delay time.
Range: 3 cmH
O to CPAP , increments of 0.5
2
Pressure
The device operating pressure may be displayed in one
Unit
of three pressure units: cmH
Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS • England • 44-138-444-6688
IntelliPAP™AutoAdjust • SleepCube™ AutoAdjust • SleepCube Plus™ CPAP
Use the
LEFT
ton and
RIGHT
button to navigate
through the menu.
Use the
UP
and
DOWN
adjust the settings.
Factory Default
English
DV54: AutoAdjust
DV53: CPAP
15.0 cmH
0, increments of 0.5
2
5.0 cmH
20 min
cmH
0
2
O, hPa or mbar.
2
1.8 m (6 ft)
N/A
N/A
N/A
N/A
English
DV54: AutoAdjust
DV53: CPAP
10.0 cmH
20 min
4.0 cmH
cmH
0
2
O, hPa or mbar.
2
DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881
Quick Set-Up Guide
Menu Item
Description
Tubing
The device is supplied with air supply tubing of 1.8 m
Length
(6 ft) in length. The device may also be operated with air
supply tubing of 3 m (10 ft). This setting should be
updated to match the tubing length for maximum
pressure accuracy.
Compliance
Cumulative hours of "while breathing" time.
Meter
Advanced
Press the  button to enter. Reference the Advanced
Menu
Submenu section for details.
Enable
Press the  button to enter. Reference the Enable
Menu
Submenu section for details.
but-
Lockout
Press the  button to enter. Reference the Lockout
Menu
Submenu section for details.
Advanced Submenu
Apnea %
The absolute value below which the respiratory flow
must fall to be considered as an apnea.
Range: 0-20%, increments of 5
button
Apnea
The number of seconds the respiratory flow must
button to
Duration
remain below the apnea percent to be detected as an
apnea.
Range: 6-150 seconds, increments of 2
Hypopnea % The percent which the respiratory flow must fall from its
normal value to be considered as a hypopnea.
Range: 30%-70%, increments of 5
Hypopnea
The number of seconds the respiratory flow must
Duration
remain below the hypopnea percent to be detected as a
hypopnea. Range: 6-150 seconds, increments of 2
Exit
Press the  button to exit and return to Clinical Menu.
Submenu
Enable Submenu
Auto-ON
0
This option controls the Auto-ON feature, which
2
automatically starts the flow of air after breathing once
or twice into the mask
0
Auto-OFF
This option controls the Auto-OFF feature, which
2
automatically stops the flow of air after taking off the
mask.
Mask Fit Chk Displays a reminder to the patient to check the fit of the
mask if airflow was greater than 95L/Min for more than
10% of the previous therapy session (including vent
leak). The previous therapy session must be greater than
3 hours or a reminder will not be displayed.
Notifications Setting to enable mask/filter check notification.
Exit
Press the  button to exit and return to Clinical Menu.
Submenu
Lockout Submenu
Delay Time
Used to allow patient access to CPAP delay time setting.
Lockout
Range: Locked, Unlocked
Tube Length
Used to allow patient access to tubing length setting.
Lockout
Range: Locked, Unlocked
Exit
Press the  button to exit and return to Clinical Menu.
Submenu
Other Settings
Filter Cleaning Reminder – Displays a reminder to the patient to clean the filter
every 50 hours of use, or approximately every 10 days. This feature is controlled by
the notifications setting and is preset to enabled.
Other setup information
• This device features automatic altitude adjustment for altitudes up to 2743m
(9000ft) and needs no manual adjustment.
0
2
• This device features a pressure sensor therefore no fine pressure adjustments are
needed.
• The SmartLink Data Management System may be used for additional reminders and
notifications.
0
2
• Clear compliance meter by pressing Down key for eight seconds with compliance
meter shown (value will go to zero).
All device settings are immediately updated in the device as you make adjustments.
When finished press the ON/OFF button once to exit the Clinical Setup Menu.
EN
Factory Default
1.8 m (6 ft)
N/A
N/A
N/A
N/A
10%
10 seconds
50%
10 seconds
N/A
Enabled
Disabled
Enabled
Enabled
N/A
Unlocked
Unlocked
N/A
©2007 DeVilbiss Healthcare 12.07
DV54D-125 Rev. A

Advertisement

loading

Summary of Contents for DeVilbiss IntelliPAP AutoAdjust

  • Page 1 Unit of three pressure units: cmH O, hPa or mbar. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS • England • 44-138-444-6688...
  • Page 2 Cuando haya terminado, pulse el botón ON/OFF una vez para salir del menú de configuración clínica. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • EE. UU. • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A...
  • Page 3 : cm H O, hPa ou mbar. touche ON/OFF pour sortir du Menu Clinique. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A...
  • Page 4 Maßeinheiten angegeben: cmH O, hPa oder mbar. Einrichtungsmenü zu schließen. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS • England • 44-138-444-6688...
  • Page 5 Menu di impostazioni cliniche. O, hPa o mbar. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A...
  • Page 6 Alle apparaatinstellingen worden onmiddellijk bijgewerkt in het apparaat terwijl u aanpassingen uitvoert. Druk eenmaal op de knop ON/OFF om het menu Klinische instellingen te verlaten. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A...
  • Page 7 ON/OFF för att lämna menyn Clinical Setup. tryckenheter: cm H O, hPa eller mbar. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A...
  • Page 8 Kaikki laitteen asetukset päivitetään välittömästi säätöjä tehtäessä. Kun säädöt on tehty, poistu kliinisten asetusten valikosta painamalla VIRTA-painiketta yhden kerran. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS • England • 44-138-444-6688...
  • Page 9 O, hPa eller mbar. er færdig, skal du trykke én gang på afbryderknappen ON/OFF for at afslutte den kliniske opsætningsmenu. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A...
  • Page 10 Verdiområde: 3 cmH 0 til CPAP , inkrementer på 0,5 oppsett-menyen. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A Sunrise Medical Ltd. • Sunrise Business Park • High Street • Wollaston, West Midlands • DY8 4PS • England • 44-138-444-6688...
  • Page 11 ρυθμίσεις. Όταν τελειώσετε, πατήστε το κουμπί ON/OFF μια φορά για να βγείτε από το Μενού CPAP, διαστήματα των 0,5 Κλινικής Εγκατάστασης. DeVilbiss Healthcare • 100 DeVilbiss Drive • Somerset, PA 15501 • USA • 800-338-1988 • 814-443-4881 ©2007 DeVilbiss Healthcare 12.07 DV54D-125 Rev. A...

This manual is also suitable for: