Specific System Settings
Fig. 14: Buzzer setting
Under Configuration click <Buzzer>.
Make changes according to the following description:
Block time settings
- Warning time limit: The buzzer sound starts when the entered radiation time has
elapsed.
- Block time limit: After the entered radiation time has elapsed - without reset of the
"warning time limit" - the radiation is blocked the next time it is selected. Every reset
of "Warning Time Limit" also resets the "Block Time Limit" to "0".
Buzzer mode (indirect, I.I.)
- Buzzer off: Buzzer sound off. Exceptions: Push mode and "Warning Time Limit" has
ended.
- Buzzer on: Buzzer sound always during radiation.
The country-specific regulations must be observed for the max. fluoroscopy time and radi-
ation blockage.
Factory setting:
Siemens AG
Medical Solutions
SPR2-330.815.01.03.02
12.06