Configuration - Siemens ACUSON Sequoia Replacement

Hide thumbs Also See for ACUSON Sequoia:
Table of Contents

Advertisement

14

Configuration

Once the system has been reassembled:
1. Boot system to SUI.
2. Select UI diagnostics in the service diagnostics menu.
3. Verify user interface passes the following tests:
- Keyboard
- LCD
- LED
ACUSON Sequoia
Sequoia Elastomer Replacement
US02-109.841.01.01.02
02.04
Page 14 of 16
sd