Siemens ARCADIS Avantic Start-Up Instructions page 48

Hide thumbs Also See for ARCADIS Avantic:
Table of Contents

Advertisement

48
Tab. 6
Measurement points for the protective conductor resistance measurement
Measurement points
Measurement from the protective conductor connection of the power plug to the:
Base unit, handle attachment screw for I.I. cover
Base unit, one attachment screw at I.I. grid
Base unit, one attachment screw at I.I. flange
Base unit, one attachment screw at I.I. cap
Base unit, one attachment screw at POWERPHOS
Base unit, one screw at cover of horizontal carriage
Base unit, one lateral screw at brake pedal (left or right)
Base unit, back cover (electronics box)
Base unit, equipotential bonding connection
Base unit, near footswitch cable outlet (electronics box)
Base unit, one lifting column screw at the base plate of the electronics box
Base unit, base plate of electronics box (at unpainted location)
Base unit, one screw at the back of each monitor
Monitor cart, one screw at monitor support
Monitor cart, one screw at the small, top back cover
Monitor cart, one screw at the large, lower back cover
Monitor cart, one screw at back cover of switch-on assembly
Monitor cart, one screw at the front cover
Monitor cart, one screw at the right side cover
Monitor cart, one screw at the left side cover
Monitor cart, one screw at the right handle
Monitor cart, one screw at the left handle
ARCADIS Avantic
SPR2-330.815.01.03.02
12.06
Final Work Steps
Page 48 of 66
CS PS SP
Siemens AG
Medical Solutions

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents