Item
Parameter
1
Temp(ºC)
2
Time
3
Ramp(C/S)
4
+Temp
5
+Time
Cycle1
×00
6
From 00
to 00
7
Grad.
8
Date
9
Time
10
Default File
11
Default User
12
Control Mode
13
Sample Volume
14
Hotlid Temp
15
Block Temp
16
Key Beep
17
Run End Beep
APPENDIX 2 Parameter Values
Scale
Temp. Control point
Time of Temp.
Control(mm:ss)
Ramping Rate
(ºC/Second)
Temp. Gain of Every
Cycler
Time Gain of Every
Cycler(mm:ss)
Section Numbers
Cycler Numbers from
beginning to the end
The start segment of
the section
The end segment of
the section.
Gradient Temp.
Difference( )
Date(YY-MM-DD)
Time(HH:MM:SS)
The default file name
The default user
name
Control Mode
Sample Volume(ul)
State of Hotlid: OFF
State of Hotlid:
Open( )
State of Block Temp
during the heating of
Hotlid(ºC)
Key Sound (ON or
OFF)
Run End Sound (ON
or OFF)
Range
4.0~99.0
--:--
(set the value before
":"through "Function"
key.)
00:00~59:59
0.1 ~5.0
#.#
0.0~9.9
0:00~9:59
1~5
1~99
X(and existed segment
no.)
≤ X+15
0 ~30
00(~99)-01(~12)-
01(~31)
00:00:00~23:59:59
Max. length is 11
characters.
Max. length is 11
characters.
Block
Tube
10~200
OFF
30~99
0~99
OFF/ON
OFF/ON
18
Note
The time of Temp. control is
infinite, meanwhile, the hotlid
turned off.
Maximum of Ramping Rate
Cycle1:×00 From 00 to 00
year-month-date
hour:minute:second
Under
this
mode
reaching
target
temperature
namely
entering
into
thermostable state), the time
for
reagent's
temperature
under target temperature is
less than setting time (see Fig.
1).
Under
this
mode
entering into the thermostable
state
after
reaching
temperature and decreasing
about
3°C),
reagent's
temperature can be close to
target temperature very fast
(see Fig. 2).
Humming
Humming continuously
(once
the
(once
target
Need help?
Do you have a question about the MultiGene Mini and is the answer not in the manual?