2 - 8
Fig. 15
Cables within the energy chain
•
Remove the plate (4/Fig. 15).
•
Remove the fabric tape from both ends of the energy chain.
•
Open the energy chain fully (1/Fig.15).
•
Remove the Plexiglas cover (3/Fig. 15).
•
Release the frame of the PC boards D1 and D30 on the opposite side of the basic unit,
and pull it out to the side (Fig. 16).
•
Remove the ferrite core from the M2 X1 cable (5/Fig. 15).
•
Trace the cables inside the energy chain, and loosen the corresponding cable ties,
cable clamps and plug/screw connections in the basic unit.
•
Pull any cables back towards the energy chain to facilitate threading the cables in and
out.
POWERMOBIL
Fig. 16
SPR2-240.841.01
Rev. 01
05.03
Disassembly
Page 8 of 12
CS PS 24
2
Siemens AG
Medical Solutions