Fig. 17

POWERPHOS

Run the cables in the POWERPHOS C-arm segment so that the cables are not crushed
when inserting the POWERPHOS.
Run the loose contacts to the plug Z66 X1, and reinsert them (Fig. 17).
Tie the cables using cable ties.
Reinsert the POWERPHOS in the C-arm segment. Make sure that the cables are not
crushed.
t
Screw the two inverter connectors back on. Tighten the nuts with a torque of 4.8 Nm.
Fasten the protective ground wires.
Connect all plugs to the POWERPHOS.
Connect the collimator plug.
Attach the sterile cover above the collimator.
If applicable, assemble the DIAMENTOR dose measuring chamber with insulating disk
and attach it.
Attach the two cable clamps (1/Fig. 18). Make sure that the cable shields makes good
ground contact.
Fit the fan assembly, and connect it observing the connection order as previously noted.
Tie or attach all cables using cable ties.
Fasten the POWERPHOS to the C-arm. Make sure that the screws are tightened evenly.
Siemens AG
Medical Solutions
Fig. 18
SPR2-240.841.01
Rev. 01
05.03
Page 7 of 10
CS PS 24
3 - 7
3