Toothed belt and concealed cable run in the C-arm

The C-arm position is in 0° orbital (center position) and 180° angulation (X-ray image in-
tensifier below, tube on top, one turn of the cable around the angulation axis,
see 1/ Fig. 11).
Release the two tension screws on the image intensifier side (1/Fig. 10); the tension on
the toothed belt is released.
Unscrew the two mounting screws (2/Fig. 10) of the clamping plate.
The toothed belt is released.
Loosen the two screws (1/Fig. 10), and remove the clamping piece (3/Fig. 10).
Pull the C-arm cable out of the C-arm.
Release the two tube-side screws (1/Fig. 9), and pull the C-arm cable out of the C-arm.
POWERMOBIL
Fig. 10
SPR2-240.841.01
Rev. 01
05.03
Disassembly
Page 6 of 12