Download Print this page
Advantech PCLD-8811 Startup Manual
Advantech PCLD-8811 Startup Manual

Advantech PCLD-8811 Startup Manual

Low-pass active filter board

Advertisement

Quick Links

PCLD-8811
Low-Pass Active Filter Board
Packing List
Before installation, please make sure that you have:
• 1 x Filter Board
• 2 x Screws
• 1 x Pair of Tweezers
If anything is missing or damaged, contact your
distributor or sales representative immediately.
Features
• The low-pass active filter has the flattest frequency
response as possible in the passband
• 8 set of low-pass Frequencies form 10Hz up to 40KHz
• 2 single-ended or 1 differential AI inputs
• Offset and gain calibration
General Specifications
General
• Connectors
2 x 5 pin male Connectors
• Dimensions
80 x 56 mm
• Filter Type
Active Filter - Butterworth
• Offset Error
± 1 LSB ( LSB resolution refer to
taken DAQ spec. )
• Gain Error
± 1 LSB ( LSB resolution refer to
taken DAQ spec. )
• Filter Frequency
-3dB,10Hz, 50Hz,100Hz, 500Hz
1KHz, 5KHz, 10KHz,40KHz
• Max. Input Voltage ± 10 V
• Input Impedance
1G Ω / 2pF
• Operating Temp.
0 ~ 60°C (32 ~ 140°F)
• Storage Temp
-20~ 70°C (-4 ~ 158°F)
Notes
For more information on this and other Advantech
products, please visit our websites at:
http://www.advantech.com
For technical support and service:
http://www.advantech.com/support/
This startup manual is for PCL-8811.
Part No: 2003881100
1 Startup Manual
Getting Started
Before you can use the PCLD-8811, you must enable
filter function on PCLD-8810 (please refer to PCLD-
8810 Startup manual) and you can refer to the below
table to set up the cut-off frequency.
Diagram
1. Plug the PCLD-8811 onto the PCLD-8810
2. Screw the pillars to the cylinders.
Note: Ensure that the pillar is aligned to the cylinder.
1st Edition
May 2015

Advertisement

loading
Need help?

Need help?

Do you have a question about the PCLD-8811 and is the answer not in the manual?

Questions and answers

Summary of Contents for Advantech PCLD-8811

  • Page 1 Packing List Getting Started Before installation, please make sure that you have: Before you can use the PCLD-8811, you must enable filter function on PCLD-8810 (please refer to PCLD- • 1 x Filter Board 8810 Startup manual) and you can refer to the below •...
  • Page 2 Gain Calibration Calibration Process After offset calibration finished, please execute gain Before you can use PCLD-8811 to acquire data cor- calibration as follow: rectly, you must have offset and gain calibration. 1. Click Device Setting and Setting Connect Disconnect PCLD-8810 and open the DAQNavi to Type and Value Range Type of AI Chan- view the Device as Figure 1 .