Care And Handling; Maintenance & Service; Environmental Conditions - PRO Scientific Bio-Gen PRO200 Operating Manual

Homogenizer
Table of Contents

Advertisement

RUNNING A GENERATOR OR CHAMBER ASSEMBLY WITHOUT LIQUID MEDIA CAN
CAUSE DAMAGE TO THE BEARINGS, UNLESS THE LOWER BEARING OF THE
GENERATOR HAS BEEN REPLACED WITH A SEALED AND SHIELDED STAINLESS
STEEL BEARING.
WHEN USING THE MULTI-GEN 7XL PROBES IT IS NOT RECOMMENDED RUNNING
AT MORE THAN HALF SPEED. (16,000RPM)
4) Turn the On/Off switch to the ON position.
5) The speed is controlled by the slide switch on the face of the motor unit. The PRO200 has a gradually increasing
speed control which relates approximately to the following segments:
5,000 to 8,000 rpm
9,000 to 11,000 rpm
12,000 to 17,000 rpm
18,000 to 24,000 rpm
25,000 to 35,000 rpm (115v) or 25,000 to 33,000 rpm (220v)

Care and Handling:

Please unpack the apparatus carefully and check that it is not damaged. It is important that any damage that
occurred in transport is detected at the time of unpacking. If you do find such damage, the carrier must be notified
immediately.
Maintenance & Service:
1. The homogenizer should be given the care normally required for any electrical appliance.
2. Avoid wetting or unnecessary exposure to fumes.
3. The finish can be washed with water and soap or detergents, using a cloth or sponge.
a. Do not allow water to get inside the unit.
b. Allow drying before using it.
4. When necessary to replace motor unit brushes;
a. Remove pair of black caps from either side of Bio-Gen PRO200 motor unit
b. Remove brush assembly
c. Replace with p/n 01-31292 (must replace entire pair)
d. Replace black caps
e. Run in brushes
i. Let the Bio-Gen PRO200 run at lowest speed for five minutes

Environmental Conditions:

Non-Operating Storage:
Temperature: -20 to 65 deg. C (-4 to 149 deg. F)
Humidity:
Operating Conditions:
Temperature: 18 to 33 deg. C (64 to 91 deg. F)
Humidity:
Altitude:
Installation Category II and Pollution Degree 2 in accordance with IEC 664.
7
20% to 85% RH, non-condensing
20% to 85 % RH, non-condensing
0 to 6,562 ft. (2000 M) above sea level