3)
Attach one end of the encoder pickup cable to the top of the motor unit and the other end to the digital speed
control box.
4) Make sure that the "ON/OFF" switch on the digital speed control box is in the "OFF" position.
5) Turn the speed control knob on the digital speed control box to the minimum position.
6) Turn the speed control dial on the top of the motor unit to position 6. If the speed control dial is not in the
maximum position (position 6) the motor will not run properly with the digital speed control box.
7) Turn the motor unit's on/off switch ( I / O ) to the on position.
8) Insert the female end of the power cord set into the back of the digital speed control box.
9) Insert the male end of the power cord into the proper voltage outlet.
10) After all of the assembly operations have been completed and the electrical connections checked, a test run of
the motor and digital speed control box can be made.
RUNNING A GENERATOR OR CHAMBER ASSEMBLY WITHOUT LIQUID MEDIA CAN
CAUSE DAMAGE TO THE BEARINGS, UNLESS THE LOWER BEARING OF THE
GENERATOR HAS BEEN REPLACED WITH A SEALED AND SHIELDED STAINLESS
STEEL BEARING
11) Turn the power switches on both the rear and face of the digital speed control box to the "ON" position.
12) Turn the speed control knob clockwise and the motor will start to turn and the motor speed will be displayed on
the digital display located on the face of the speed control.
a. Please note, the display shows real time motor speed.
NEVER ATTACH OR DETACH ENCODER PICKUP CABLE FROM THE MOTOR OR
SPEED CONTROL BOX WHILE EITHER ON/OFF SWITCH IS IN THE "ON" POSITION

Care and Handling:

Please unpack the apparatus carefully and check that it is not damaged. It is important that any damage that
occurred in transport is detected at the time of unpacking. If you do find such damage, the carrier must be notified
immediately.
Maintenance & Service:
1. The homogenizer should be given the care normally required for any electrical appliance.
2. Avoid wetting or unnecessary exposure to fumes.
3. The finish can be washed with water and soap or detergents, using a cloth or sponge.
a. Do not allow water to get inside the unit.
b. Allow drying before using it.
4. When necessary to replace motor unit brushes please contact a PRO Scientific Service technician (P#203-267-
4600, sales@proscientific.com) or an authorized distributor.