Environmental Conditions:
Non-Operating Storage:
Temperature: -20 to 65 deg. C (-4 to 149 deg. F)
Humidity:
Operating Conditions:
Temperature: 18 to 33 deg. C (64 to 91 deg. F)
Humidity:
Altitude:
Installation Category II and Pollution Degree 2 in accordance with IEC 664.
Safety
1) Never attempt to hold the lower end of the generator while the generator is attached to the motor.
2) Over tightening the rotor knife onto the rotor shaft can result in breaking the shaft and/or distortion of the
rotor knife.
3) Any servicing of the homogenizer motor unit, except brush replacement, should be performed by the PRO
Scientific Service Department.
4) The use of any accessories or attachments other than those supplied by the manufacturer may be hazardous
and voids all warranties.
5) The motor unit is supplied with sealed ball bearings and requires no additional lubrication. Any additional
lubrication to the motor can result in bearing and/or motor failure.
6) Running a generator or chamber assembly without liquid media can cause damage to the bearings and
consequently damage the generator.
15
20% to 85% RH, non-condensing
20% to 85 % RH, non-condensing
0 to 6,562 ft. (2000 M) above sea level