Attachment of the Multi-Gen Generator to the Adapter: (see figure 5)

1) Remove one generator from the box.
2) Push the sleeve on the adapter upward.
3) While the sleeve is pulled back, insert the grooved end of the generator
into the bottom of the adapter.
4) Once the generator has been inserted as far as possible into the bottom
of the adapter, the bottom sleeve of the adapter can be released.
Releasing the bottom sleeve will lock the generator in place. Pulling
gently on the generator will ensure that the generator is locked
securely in place.
Removal of the Multi-Gen Generator from the
Adapter:
1) With the motor in an upward position, (generator pointing downward),
slide the bottom sleeve of the adapter upward. Sliding the bottom sleeve of the adapter upward will allow the
generator to drop out of the adapter.
**CAUTION: CARE SHOULD BE TAKEN WHEN HANDLING THE MULTI-GEN 7XL, ESPECIALLY WHEN
REMOVING THE MULTI-GEN 7XL FROM THE ADAPTER. DO NOT LET THE GENERATOR DROP AS THIS

Operating the Multi-Gen Generator:

Once attached to the motor and the adapter, the Multi-Gen generators can be used like any other PRO Scientific
generator.
**CAUTION SHOULD BE TAKEN NOT TO RUN THE GENERATOR DRY, WHICH CAN RESULT IN THE FAILURE OF THE PTFE
BEARINGS AND SUBSEQUENT DAMAGE TO THE GENERATOR. DURING OPERATION OF THE UNIT, PLEASE ENSURE
THAT THE LOWER BEARING IS WITHIN THE LIQUID OF THE SAMPLE BEING PROCESSED.**
PLEASE NOTE THE SEPARATE AND DIFFERENT STEPS FOR ASSEMBLING AND
DISASSEMBLING THE MULTI-GEN 7 AND THE MULTI-GEN 7XL

Dismantling of MULTI-GEN 7 ONLY

1. Insert the blade end of the screwdriver (supplied in the motor unit tool kit) into the bottom of the generator,
locking the knife in place.
2. Continue with step 3.

Dismantling of MULTI-GEN 7XL ONLY

1. Insert the allen wrench end of the screwdriver (supplied in the motor unit tool kit) across and through the open
slot of the XL generator tube
2. Continue with step 3.
3. While firmly holding the drive collar at the upper end of the generator (opposite end of the screwdriver) turn
the screwdriver counter clockwise. If the drive collar is not held securely it will spin with the screwdriver,
thereby not allowing the knife assembly and rotor shaft to be unscrewed from the drive collar.
4. With the drive collar removed, the rotor shaft knife assembly will slide out the bottom of the generator tube.
5. The upper PTFE bearing can now be removed from the upper end of the generator tube.
6. The lower PTFE bearing can now be removed from the lower end of the generator tube.
23
COULD DAMAGE THE TEETH OF THE GENERATOR.**
***WARNING***
Figure 5