Calibration - Ecolab DG1-III-DLA-PCB Operating Instructions Manual

Microprocessor-controlled dosing unit
Table of Contents

Advertisement

Available languages

Available languages

Starting up
7.4

Calibration

Calibration is carried out by means of the potentiometer 'CAL DLA' .
Example
Set concentration: 0.50 %
Potentiometer in max position (right stop):
0.50 % + 20 % von 0.50 % = 0.50 % + 0.2 x 0.50 = 0.60 %
Potentiometer in min position (left stop):
0.50 % - 20 % von 0.50 % = 0.50 % - 0.2 x 0.50 = 0.40 %
MAN049250 Rev. 1-01.2021
By means of this potentiometer, the dosage can be adjusted by
approx. +/-20 % in relation to the set concentration.