CADD Legacy Plus 6500 Operator's Manual page 85

Ambulatory infusion pump
Table of Contents

Advertisement

Press and hold
Press and hold
and then stop. The pump will alarm and display "Reservoir Volume
Empty." Press
Starting/Stopping the Pump
Program the pump with the following values:
Reservoir Volume:
Continuous Rate:
Given:
Program the Air Detector Off (see Section 4, Biomed Functions).
Press and hold
each accompanied by a beep. A review of the programmed param-
eters then appears. The main screen should appear with "RUN" in the
display.
To stop the pump, press and hold
by 3 sets of dashes that disappear one at a time, each accompanied
by a beep. The main screen should appear with "STOPPED" in the
display.
Activation Timing Check
Reprogram the Reservoir Volume to 1.0 ml and clear the Given
screen.
Press and hold
pump should sequentially display all of the programmed values. Start
a timer at the first motor activation.
Count the activations. One activation should occur every 6 seconds.
Approximately 66 seconds and 10 activations later, the Reservoir Vol-
ume empty alarm should occur. The display should show "Reservoir
Volume empty."
until 3 series of dashes appear. Release
. The pump should prime 10 double activations
.
. "Starting" appears followed by 3 sets of dashes,
until 3 dashes disappear from the display. The
1.0 ml
50 ml/hr
0.00 (press
¤
. "Stopping" appears followed
Section 5: Reference
.
)