Section 2: Pump Setup and Programming
IDM: Stopping the Pump During the Dose
Stopping the pump while a dose is in progress will shift all subsequent
doses by the amount of time the pump is stopped.
Pump is
All
stopped for
subsequent
10 minutes
doses shift by
10 minutes
Dose
Dose
If you wish to make up for the lost time in the cycle, it is recommended
that you wait until the current dose is completed. If the pump is stopped
while a dose is infusing and you reprogram the "Dose Starts in" screen,
then the pump will not deliver the remainder of that dose. (To determine
the amount of time remaining in the current dose, use the Dose Remain-
ing screen in the programming loop.) Then stop the pump, change to LL0
or LL1, and reprogram the delay on the "Dose Starts in" screen.
Resuming the Dose
To resume delivery of a dose that has been stopped in progress, simply
restart the pump.
Resetting the Cycle
If you wish to reset the cycle, stop the pump during KVO, change to LL0
or LL1, and reprogram the start of the Next Dose.
If a dose has been stopped in progress and you wish to cancel the current
dose and reset the cycle, make sure the pump is in Lock Level 0 or Lock
Level 1. The Dose Remaining screen with the remaining hours and min-
utes of the dose will be displayed, followed by the "Dose Starts in" screen.
Reprogram the "Dose Starts in" screen, which will cancel the remainder of
the dose, and the next cycle will start after the delay you selected.